EXPMT 2000 - Pacemaker Analyzer
A compact Pacemaker Analyzer / Tester for testing all
Pacemakers including Atrial / Ventricular and Transthoracic Pacers.
Accurate, Compact & Best value.
EXPMT 2000 - Pacemaker Tester & Analyzer
EXPMT 2000,
Pacemaker Analyzer is the ultimate Pacer tester designed for evaluating
and testing of all Atrial / Ventricular Pacemakers including Transthoracic Pacers.
The New EXPMT 2000, External Pacemaker Analyzer, has evolved from the best
selling EXPMT 100 Pacer Analyzer. It will test and verify all the functions of any external pacemaker including
Transvenous, Transthoracic, and A-V Sequential models.
EXPMT 2000, Pacemaker Analyzer with its new look, new features, new controls,
and new display combine to make it the easiest to use and the most comprehensive pacer analyzer available.
The operation of this Pacemaker Analyzer is menu driven via eight tactile keys and a large four line LCD display. Test measurements are
simultaneously presented on the display. Five ECG snaps are standard along with an RS 232 serial port that allows
for printing botha heading and test measurements on a separate printer. Information may also be sent to a Palm Pilot.
The External Test Load has twelve user selectable loads from 100 to 1000 Ohms for use with transcutaneous pacers.
Features:
- Easy to Use, Fast & Accurate
- 12 Selectable Test Loads
- 5 ECG Test Leads
- RS 232 Interface and a dedicated Print Key for
simplicity
- Accurate Test results, Pulse Rate, Width,
Current (amplitude)
- Demand Sensitivity Tests, Sensed and Paced RefactoryTests
- A-V Interval, Pulse Energy
- 60 Hz/ 50 Hz Immunity Tests
Back to top
Specifications:
- Test Load: 50 Ohm fixed
- 12 Selectable step from 100 to 1000 ohms with EXPMT Load Selector
- Current: 0.1 to 25 mA
- Accuracy: ±10%, 1 to 10 mA,
±5%, 10 to 200 mA
- Pulse Rate: 30 to 800 PPM
- Accuracy: ±1% or 2 ppm
- Pulse Width: 0.5 to 80 ms
- Accuracy: ±10%, 0.5 to 5 ms,
±5%, 5 to 25 ms,
±1%, 25 to 80 ms
- Sensed and Paced Refractory Period: 25 to 500 ms
- Accuracy: ±10 ms
- 50/60 Hz Immunity Test: 0 to 10.6 mV in 0.7 mV steps
- Test Load: 50o Ohms ± 1%
- Current: 0.1 to 25 mA
- Accuracy: ±10%, 0.1 to 10 mA
±5%, 1 to 25 mA
- Pulse Rate: 30 to 800 PPM
- Accuracy: ±1% or 2 ppm
- Pulse Width: 0.5 to 80 ms
- Accuracy: ±10%, 0.5 to 5 ms
±5%, 5 to 25 ms
±1%, 25 to 80 ms
- Sensed and Paced Refractory
- Period: 25 to 500 ms
- Accuracy: ±10 ms
- 50/60 Hz Immunity Test: 0 to 102 mV in 6.8 mV steps
- Waveforms: SQR, TRI, and SSQ
- Amplitude: 0 to 100 mV.
- Width: 10, 25, 40, 100, and 200 ms
- Pacer Rate: 30 to 120 ppm
- A - V Interval: 0 to 999 ms
- Display: 4 Line LCD Display
- Housing: ABS Plastic
- Power Requirements: 9V Battery or AC Operation
- Physical Dimensions: 7.75 x 4.0 x 1.5 inches
(19.7 x 10.2 x 3.8 cm)
- Weight: 1 lbs (0.9 kg)
- Environmental:
Operating Temperature: 59° to 95° F (15° to 35° C)
Storage Temperature: 32° to 131° F (0° to 55° C)
Back to top
Ordering Information:
| Part Number |
Description |
Price |
| 950 |
Pacemaker
Analyzer (Include Carrying case, Ac Adapter,
Interface cables) |
$2150
(Trade in old) |
| 950-Case |
Carrying
case |
Call |
| 950-302 |
Ac
Adapter |
Call |
| 950-Interface-Cables |
Universal
Interface cable |
Call |
| Specific
Interface |
Call
1-800-547-6557
for