Defibrillator Analyzers
Electrical Safety Analyzers
Gas Flow Meters
Laser Meters
Light Meters
NIBP Simulators
Pacemaker Analyzers
Patient Simulators
Pressure Meters
Sound Level Meter
Tachometers
Ultrasound Power Meters



Electrical Safety Tester/12 Lead ECG Simulator
A Microprocessor based, full Function Electrical Safety Analyzer with a buit-in 12 lead ECG Simulator


MULTI-PRO 2000 - A Full Function Electrical Safety Analyzer. with 12 lead ECG
The MULTI-PRO 2000, Electrical Safety Tester is a micro-processor based, full function electrical safety analyzer with a built-in 12 lead ECG Simulator. The combination instrument is designed to test the electrical safety of all types of medical equipment by hospital and field service biomedical technicians. In addition, the built-in ECG Simulator allows performance testing of ECG equipment.

The digital circuitry of the compact MULTI-PRO 2000, Electrical Safety Tester eliminates all mechanical switching and delivers in matched testing accuracy. The menu driven soft keys provide fast and easy operation. Measurements results and menus are presented on the large LCD display.

The MULTI-PRO 2000, Electrical Safety Tester tests wall outlet polarity and measures line voltage, instrument current, chassis resistance, chassis leakage current, and leads leakage current. It also performs lead isolation testing and point to point testing. Calibrated test points are available for verifying its performance.

The ECG Simulator displays a normal sinus rhythm waveform with 13 amplitude and 14 rate selections. Square, Sine, and Triangle waveforms for performance testing are available with 14 frequency and 13 amplitude selections.

Optional software is available to store measurements and to interface to a Palm Pilot or independent printer.

The MULTI-PRO 2000, Electrical Safety Tester is a light weight, versatile tool that easily performs electrical safety testing to the requirements of both in-house and field biomedical service technicians. 


Features:
  • Micro Processor Controlled

  • Menu Operated

  • Compact, Lightweight, Rugged

  • Accurate, and Reliable

  • 12 Lead ECG and PErformance Waveforms

  • 20 Ampere Rating

  • Quick and Easy to Operate

  • Kelvin Cable


Measurements:
  • Receptacle Polarity

  • Line Voltage

  • Chassis Resistance

  • Leakage Current

  • Leads Leakage

  • Isolation Current

  • Point to Point Testing
Back to top
Specifications:
    Technical

    Line Voltage:-
       Range: 1 to 300 V 
       Accuracy: ± 2% + 1 V 

    Leakage Current:-
       Range: 0 to 2500 µA autoranging 
       Accuracy:-
           DC to 100 KHz: ± 2% FS ± 1 Digit 
           100 KHz to 1 MHz: ± 4% FS ± 1 Digit 

    Resistance:-
       Range: 0 to 2500 mW, auto ranging 
       Accuracy: ± 2% FS ± 1 Digit 

    Device Current:-
       Range: 0 to 20 A 
       Accuracy: ± 2% FS ± 1 Digit 

    Controls:-
       On Off Power Switch
       Sixteen Tactile Feel Keys
       Ten Color Coded ECG Jacks
       Two Input Jacks
       RS 232 Serial Port connection (Optional)

    ECG Simulator:-
       12 Lead Normal Sinus Rhythm and Pulse Waveforms:-
           Amplitude: 13 selectable steps 
                             0.1, 0.2, 0.3, 0.4, 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0  
                             and 4.5 mV 
           Rate: 14 selectable steps
                     30, 60, 70, 80, 90, 100, 120, 150, 180, 210, 240, 270, 300, 
                     and 350 BPM 

       Square, Triangle, and Sine Performance Waveforms:-
          Amplitude: 13 selectable steps
                            0.1, 0.2, 0.3, 0.4, 0.5, 2.0, 2.5, 3.0, 3.5, 4.0 and 4.5 mV 
          Rate: 25 selectable steps
                   0.1, 0.2, 0.3, 0.4, 0.5, 1.0, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, 
                   30, 35, 40, 45, 50, 55 and 60 Hz 


    General

    Display: 2 Line 48 character Display

    Physical Dimensions: 10.25 x 6.25 x 2.5 inches
                                     (26.00x 15.9 x 6.4 cm)

    Weight: 4 lbs (1.8 kg)

    Power Requirements: 110 VAC 50-60 Hz, 20 A
                                       220 VAC 50-60 Hz, 10A 

    Environmental:-
       Operating Temperature:59º to 95º F (15º to 35º C) 
       Storage Temperature: 32º to 122º F (0º to 50º C)

Back to top
Ordering Information:

Part Number Description Price
550 MULTI-PRO 2000 Call
553 Kelvin Cable Call
550 - Case Soft Carrying Case Call
503 Test Cable Call

Need technical help or price quote please call 1-800-547-6557 or E-mail
 